• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: A.I.D.D LONGFORD MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

A.I.D.D LONGFORD MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM Back to Search Results
Model Number 03P6834
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer observed falsely elevated magnesium results generated on the architect c4000 processing module for multiple samples.The following example results were provided: on (b)(6) 2022.Initial result = 7.9mgdl, repeat result on another analyzer= 1.3mg/dl.On (b)(6) 2022.Initial result = 6mg/dl, repeat result on another analyzer= 2.1mg/dl reference (normal) range: 1.6 to 2.6 mg/dl.No impact to patient management was reported.
 
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
 
Additional Manufacturer Narrative
The complaint investigation for false elevated architect magnesium results included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, device history record review, and field data review.Return testing was not completed as returns were not available.A review of tracking and trending did not identify any trends for the complaint issue.Labeling was reviewed and found to adequately address the issue.Data from customers worldwide was used to assess the performance of the architect magnesium assay.The median patient result for lot 39508ud00 is within established limits and comparable with other lots in the field, which indicates acceptable product performance.Device history record review on lot 39508ud00 did not show any potential non-conformances, deviations, or non-conformances associated with the complaint issue.Based on the investigation, no systemic issue or deficiency of the architect magnesium assay for lot number 39508ud00 was identified.All available patient information was included.Additional patient details are not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNESIUM
Common Device Name
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key16059207
Report Number3005094123-2022-00309
Device Sequence Number14779173
Product Code JGJ
UDI-Device Identifier00380740169862
UDI-Public00380740169862
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/15/2023
Device Model Number03P6834
Device Catalogue Number03P68-34
Device Lot Number39508UD00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/07/2022
Supplement Date Received by Manufacturer02/15/2023
Initial Report FDA Received Date12/28/2022
Supplement Report FDA Received Date02/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/14/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC C4 PROC MOD, 02P24-01, (B)(4).; ARC C4 PROC MOD, 02P24-01, (B)(4).
Patient SexUnknown
-
-