| Catalog Number |
114745L |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during the priming of a revaclear 300 , a saline solution leakage was observed.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The actual device was not available; however, two photographs of the sample were provided for evaluation.The first photo showed a dialyzer connected to tubing that is used for patient therapy.Fluid was observed outside of the device and the tubing set.The second photo showed a dialyzer that looked primed.The photo showed the hansen port potentially being damaged.The reported condition was verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|