MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY EXTENDED VISION IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY EXTENDED VISION IOL; INTRAOCULAR LENS Back to Search Results
Model Number DFT015
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Vitreous Loss (2142); Capsular Bag Tear (2639)
Date of Event 11/22/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A healthcare professional reported that after lens was introduced, it came with defective/broken haptics.The patient also had rupture of the posterior capsule with vitreous loss.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The lens was returned positioned incorrectly in the lens case.Solution was dried on the lens.The reported "broken haptic" complaint was not observed.Both haptics area attached and undamaged.The optic was cut from edge past center, typical of insertion and removal.Complaint history and product history records were reviewed and the documentation indicated the product met release criteria.The file indicated the use of qualified viscoelastics.The associated cartridge and delivery system/handpiece was not provided.It is unknown if qualified products were used.Other lens damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company¿s release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The reported "rupture of the posterior capsule with vitreous loss" could not be confirmed.Replacement lens information was not provided.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY EXTENDED VISION IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16059611
Report Number1119421-2022-02751
Device Sequence Number6382484
Product Code POE
UDI-Device Identifier00380652421195
UDI-Public00380652421195
Combination Product (Y/N)N
Initial Reporter CountryPM
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 02/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/21/2026
Device Model NumberDFT015
Device Lot Number15274216
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/06/2022
Supplement Date Received by Manufacturer01/23/2023
Initial Report FDA Received Date12/28/2022
Supplement Report FDA Received Date02/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CENTURION ACTIVE SENTRYHAND PC; PROVISC OVD; VISCOAT OVD
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-