| Model Number |
VANC3 |
| Medical Device Problem Code |
Low Readings (2460)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation determined that an electronic file did not contain the vanc3 application.The missing information might in very rare cases lead to inaccurately low vancomycin results.An updated electronic file was published to solve the issue.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for an unspecified number of samples tested with the vanc3 (online tdm vancomycin gen.3) assay on a cobas c 503 analytical unit, serial number (b)(4).No specific data on patient results being affected were provided.No questionable result was reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
The component code has been updated.
|
| |
|
Search Alerts/Recalls
|