MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1712K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number MMT-1712K |
Device Problem
Insufficient Information (3190)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/08/2022 |
Event Type
malfunction
|
Event Description
|
Information received by medtronic indicated that the customer had pump error.Troubleshooting was unknown.No harm requiring medical intervention was reported.Customer will discontinue using the device.The device will be returned for analysis.
|
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the ngp 640g insulin pump which is not marketed in the united states.However, the device is similar to the ngp insulin pump, which is marketed in the united states.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
The pump passed the active current test, sleep current test, and self test.The pump failed the rewind test due to pump error 38.Unable to perform the displacement test, prime/seating test, basic occlusion test, force sensor test, and occlusion test due to pump error 38.Successfully downloaded history files and traces using thus.Pump error 38 was noted during testing and in the downloaded pump history on (b)(6) 2022 at 11:21:33.000.No battery related alarms or alerts were noted during testing or in the downloaded pump history.The pump was cut open to perform visual inspection and found moisture damage on the pcba 1 and dried insulin on the motor.Test p-cap and reservoir locked properly into reservoir compartment during testing.The following were noted during visual inspection: pillowing keypad overlay, scratched case, and label damage (faded).Unable to confirm unspecified alarm.Pump error 38 was confirmed during testing and in the downloaded pump history due to dried insulin on the motor.Pump exposed to moisture confirmed.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|