• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® FOLEY CATHETER SILICONE COATED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® FOLEY CATHETER SILICONE COATED Back to Search Results
Model Number 265716
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the customer was trying to use a silicone elastomer foley catheter, and the inflation port came off while trying to inflate the catheter.
 
Event or Problem Description
It was reported that the customer was trying to use a silicone elastomer foley catheter, and the inflation port came off while trying to inflate the catheter.
 
Additional Manufacturer Narrative
The reported event was confirmed and the cause was unknown.A potential root cause for this failure mode could be, ¿rough shaft/inflation funnel weight out of specification/mishandling catheter.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "foley catheter caution: this product contains natural rubber latex which may cause allergic reactions.Pk6194321 4/03 rev.0 sterile unless package is opened or damaged.Do not use if package is opened or damaged.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage latex and may cause balloon to burst.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations.Bard is a registered trademark of c.R.Bard, inc.Or an affiliate.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.® single patient use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer syringe.Do not use needle.Visually inspect the product for any imperfections or surface deterioration prior to use.Recommended inflation capacities 3cc balloon: use 5cc sterile water 5cc balloon: use 10cc sterile water 15cc balloon: use 20cc sterile water 20cc balloon: use 25cc sterile water 30cc balloon: use 35cc sterile water 40cc balloon: use 45cc sterile water 75cc balloon: use 80cc sterile water do not exceed recommended capacities.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact an adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® FOLEY CATHETER SILICONE COATED
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key16060683
Report Number1018233-2022-09893
Device Sequence Number20516660
Product Code EZC
UDI-Device Identifier00801741039591
UDI-Public(01)00801741039591
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/28/2026
Device Model Number265716
Device Catalogue Number265716
Device Lot NumberMYFW3802
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/12/2022
Supplement Date Received by Manufacturer02/07/2023
Initial Report FDA Received Date12/28/2022
Supplement Report FDA Received Date02/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-