• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER KNEE CREATIONS, INC. ACCUPORT SIDE-DEL 11GA X 120MM SELF-DRILLING; FILLER, BONE VOID, CALCIUM COMPOUND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER KNEE CREATIONS, INC. ACCUPORT SIDE-DEL 11GA X 120MM SELF-DRILLING; FILLER, BONE VOID, CALCIUM COMPOUND Back to Search Results
Catalog Number 307.032
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 11/29/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: item#: 414.502; lot#: 209505622.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that during a procedure, the tip of the delivery cannula broke off in the femur and was retained by the patient.There have been no reported patient consequences as a result of the event.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; d4; d9; g3; g6; h1; h2; h3; h4; h6 visual examination of the provided photograph identified that the cannula tip is broken and the fractured off piece is not pictured.No product was returned as the product was discarded by the hospital.Medical records were not provided.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCUPORT SIDE-DEL 11GA X 120MM SELF-DRILLING
Common Device Name
FILLER, BONE VOID, CALCIUM COMPOUND
Manufacturer (Section D)
ZIMMER KNEE CREATIONS, INC.
841 springdale drive
exton PA 19341
Manufacturer (Section G)
ZIMMER KNEE CREATIONS, INC.
841 springdale drive
exton PA 19341
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16060697
Report Number3008812173-2022-00006
Device Sequence Number6639720
Product Code MQV
UDI-Device Identifier00889024406438
UDI-Public(01)00889024406438(17)250428(10)45919
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K190814
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/28/2025
Device Catalogue Number307.032
Device Lot Number45919
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2022
Supplement Date Received by Manufacturer12/30/2022
Initial Report FDA Received Date12/28/2022
Supplement Report FDA Received Date01/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-