Model Number 101/860/085CZ
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that during the use of the product, the suction line connector got detached.No patient injury reported.
 
Additional Manufacturer Narrative
One used decontaminated sample was received for investigation without its original packaging.The returned sample was visually inspected at a distance of 12 to 16 inches and normal conditions of illumination.It was identified that the suction line connector was detached, and the cover was in a broken condition confirming the complaint.Based on the analysis conducted in the sample provided the occurrence of this failure condition could be caused by the tube not submerged properly in solvent dispenser or dispenser with low level of solvent.A device history record (dhr) reviewed showed there were no issues or non-conformances reported during the manufacture of this lot number.Notification was made to production personnel as awareness to failure mode reported.The manufacturer will continue monitoring this failure condition in this product for escalation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TUBES BLUSELECT
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
7-15-26 fukushima
minneapolis, MN 55442
MDR Report Key16061103
Report Number3012307300-2022-28207
Device Sequence Number16646430
Product Code BTO
UDI-Device Identifier15019517076165
UDI-Public15019517076165
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K173384
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number101/860/085CZ
Device Catalogue Number101/860/085CZ
Was Device Available for Evaluation? No
Date Returned to Manufacturer12/01/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2022
Supplement Date Received by Manufacturer04/02/2023
Initial Report FDA Received Date12/28/2022
Supplement Report FDA Received Date04/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1