Model Number 3125-1180S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Abscess (1690); Necrosis (1971); Osteomyelitis (4533)
Date of Event 11/24/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.A review of the device history is not possible because the lot number was not communicated.Should additional information become available, it will be provided in a supplemental report.Device remains implanted in patient.
 
Event or Problem Description
As reported: "i take care of a patient after osteosynthesis of the right femur.The operation was carried out in (b)(6) 2019.The patient came to our department for femoral head necrosis with osteomyelitis and soft tissue abscess of the thigh.We need to extract the implants.This is a stryker 3 gamma nail system.The operation is scheduled for (b)(6) 2022.".
 
Event or Problem Description
As reported: "i take care of a patient after osteosynthesis of the right femur.The operation was carried out in (b)(6) 2019.The patient came to our department for femoral head necrosis with osteomyelitis and soft tissue abscess of the thigh.We need to extract the implants.This is a stryker 3 gamma nail system.The operation is scheduled for (b)(6) 2022.".
 
Additional Manufacturer Narrative
The received x-ray was reviewed and the reported necrosis can be confirmed.Only based on this picture it is not possible to determine if the necrosis is in relation with the implants, especially as the devices were implanted for more than 3 years.A device inspection was not possible since the affected device was not returned, therefore no further evaluation is possible.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONG NAIL KIT R1.5, TI, RIGHT GAMMA3® Ø11X420MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16061209
Report Number0009610622-2022-00605
Device Sequence Number12445784
Product Code HSB
UDI-Device Identifier07613252273868
UDI-Public07613252273868
Combination Product (Y/N)N
Initial Reporter CountryEZ
PMA/510(K) Number
K200869
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date09/30/2023
Device Model Number3125-1180S
Device Catalogue Number34251420S
Device Lot NumberK05BE15
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/07/2022
Supplement Date Received by Manufacturer02/27/2023
Initial Report FDA Received Date12/28/2022
Supplement Report FDA Received Date03/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/12/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-