DEPUY IRELAND - 9616671 ATTUNE PS FB INSRT SZ 2 6MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Model Number 1516-40-206
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # : (b)(4).Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported a hair was found inside the inner product.No adverse affects on the patient.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device was not returned to depuy synthes for evaluation, however photos were provided for review.Review of the provided photos reveled a hair adhered on the cellophane.Observed foreign matter does not affect sterility of the implant or compromises its sterility.The reported condition was not confirmed due to the hair was found outside of the sterile packaging and it does not affects its integrity.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: a manufacturing record evaluation were performed for the finished device dhr 151640206 / jw4604, it was manufactured on 04-april-2022.12 parts were manufactured per specification and all raw materials met specification, and no non-conformances were identified.1) quantity manufactured: (b)(4) parts.2) date of manufacture: 04-april-2022.3) any anomalies or deviations identified in dhr: no.4) expiry date: 31-mar-2027.5) ifu reference: ifu-0902-00-828.Device history review
=
> a manufacturing record evaluation was performed for the finished device (151640206/ jw4604) product and lot numbers,  and no non-conformances were identified.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was returned for analysis.Visual analysis revealed that the box packing of the attune ps fb insrt sz 2 6mm was not received for evaluation, therefore the report allegation cannot be confirmed.The observed condition not affect sterility of the implant or compromises its sterility.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation were performed for the finished device dhr 151640206 / jw4604, it was manufactured on 04-april-2022.12 parts were manufactured per specification and all raw materials met specification, and no non-conformances were identified.Quantity manufactured: (b)(4) parts.Date of manufacture: 04-april-2022.Any anomalies or deviations identified in dhr: no.Expiry date: 31-mar-2027.Ifu reference: ifu-0902-00-828.Device history review: a manufacturing record evaluation was performed for the finished device (151640206/ jw4604) product and lot numbers, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FB INSRT SZ 2 6MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY RAYNHAM, A DIV. OF DEPUY ORTHO 1219655
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key16061259
Report Number1818910-2022-26189
Device Sequence Number15113267
Product Code JWH
UDI-Device Identifier10603295049517
UDI-Public10603295049517
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K111433
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 12/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1516-40-206
Device Catalogue Number151640206
Device Lot NumberJW4604
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2022
Supplement Date Received by Manufacturer03/09/2023
04/24/2023
05/22/2023
Initial Report FDA Received Date12/28/2022
Supplement Report FDA Received Date03/09/2023
04/26/2023
05/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE PS FB INSRT SZ 2 6MM
Patient SexUnknown
-
-