| Model Number |
61238BX |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned to olympus for evaluation.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
| |
|
Event or Problem Description
|
|
Olympus received a medwatch report (uf# (b)(4)) through the mail which stated: "facilities have identified an issue with the disposable olympia uropass access sheath tips- it is breaking apart within the package and during the procedure.Multiple lot# and product # are affected.Multiple tips across various procedures broke off of the device during the procedures.The surgeons were able to recover them without any harm to the patient.Our clinicians have also noted that the tip of the device can be easily broken off within the package.Multiple lot numbers/product numbers affected".Upon further inquiry, it was reported the tips broke off within patients during procedure.Imaging was conducted and the surgery was delayed to invasively remove the tips from patient(s).No patient harm was reported and no patient information has been provided.The following medwatch reports are related.The patient identifiers are as follows: (b)(6), model #: 61224bx, lot #: 09e2100062; (b)(6), model #: 61134bx, lot #: 09j1800111; (b)(6), model #: 61238bx, lot #: 09b1900153; (b)(6), model #: 61338bx, lot #: 09a1900139; (b)(6), model #: 61324bx, lot #: 09d2000303; (b)(6), model #: 61138bx, lot #: 09l2100130.This medwatch report is for patient identifier (b)(6).
|
| |
|
Event or Problem Description
|
|
(b)(6) model# 61154bx / lot # 09j1800111.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and evaluation with correction to the initial with information inadvertently left out.Additionally, to provide a correction to the initial (b5) and to provide updates to the initial (d9 and h3).A recall (recall # 0426) of devices unexpired devices manufactured prior to 31dec2019.Recall number is unavailable at the time of this report.The reported complaint was confirmed through evaluation and investigation of the device.There is no concomitant device for this complaint due to the device not being used and the customer returning this product out of caution.The customer returned 4 devices with the lot listed in this complaint (b)(4).The devices were evaluated and there was one confirmed breakage.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is possible the suggested event occurred due to the following: 1.Uropass dilator tip breakage is proposed to be mainly caused by inconsistent manufacture process of polyethylene (pe) dilator, lot-by-lot.2.The unfavorable oxidation of polyethylene dilator material is proposed to be occurred during dilator manufacture thermal process cycle, in which the thermos-oxidation could cause brittle fracture of uropass dilator tip materials for special lots.3.Generally speaking, photooxidation of polyethylene or uropass dilator material could occur during the whole products cycle including sterilization, and shelf life.In other words, it will have affecting on uropass products for every lots but not for special lots.4.Uropass dilator tip buckling testing should be very helpful for supplier to check the quality of uropass dilator tip for each lot to make sure if it meets osta specifications or not.However, the root cause of the device malfunction could not be determined.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|