• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. (AF-ST. PAUL) 4 LESION NT2000¿ PAIN MANAGEMENT RF GENERATOR; GENERATOR, LESION, RADIOFREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. (AF-ST. PAUL) 4 LESION NT2000¿ PAIN MANAGEMENT RF GENERATOR; GENERATOR, LESION, RADIOFREQUENCY Back to Search Results
Model Number RFG-NT-2000
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
During the procedure, the generator made a continuous beep and the screen went black.The generator was restarted, however, the issue was not resolved.As a result, the procedure could not be completed.
 
Additional Manufacturer Narrative
During processing of this complaint, attempts were made to obtain complete physician information.The reported issue was confirmed.Based on the information provided to abbott and the investigation performed, the root cause of the field reported event was isolated to physical damage to the upper assembly.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.No capa is required; the root cause of the reported issue could not be confirmed to be a manufacturing, design, or quality system related occurrence.The reported issue will continue to be monitored for trends and reviewed in qdr (quality data review).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4 LESION NT2000¿ PAIN MANAGEMENT RF GENERATOR
Common Device Name
GENERATOR, LESION, RADIOFREQUENCY
Manufacturer (Section D)
ST. JUDE MEDICAL, INC. (AF-ST. PAUL)
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC. (AF-ST. PAUL)
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key16062000
Report Number2184149-2022-00317
Device Sequence Number17818930
Product Code GXD
UDI-Device Identifier05415067022417
UDI-Public05415067022417
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K111576
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRFG-NT-2000
Device Catalogue NumberRFG-NT-2000
Device Lot Number6812478
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2022
Supplement Date Received by Manufacturer04/19/2023
Initial Report FDA Received Date12/28/2022
Supplement Report FDA Received Date04/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/17/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-