DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, FRENCH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-62
Medical Device Problem Code Device Contamination with Body Fluid (2317)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/25/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a balloon broke during use on a patient causing blood to go inside the cs300 intra-aortic balloon pump (iabp).
 
Additional Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate the iabp and was able to reproduce the reported issue.The purge valve, blood detect tubing and safety disk were replaced.The fse completed all safety, functionality and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.This report is being submitted as the result of a retrospective review conducted in capa (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, FRENCH, 110V
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16062274
Report Number2249723-2022-03371
Device Sequence Number6733173
Product Code DSP
UDI-Device Identifier10607567108292
UDI-Public10607567108292
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K063525
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0998-00-3023-62
Device Catalogue Number0998-00-3023-62
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2021
Initial Report FDA Received Date12/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/24/2011
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MAQUET BALLOON CATHETER
Patient SexUnknown
-
-