DEPUY SPINE INC VIPER 2 SYSTEM ROD, STRAIGHT 480MM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY
|
Back to Search Results |
|
| Model Number |
186789480 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Vertebral Fracture (4520)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: procode: additional device product codes: osh, kwp, mnh, nkb, mni.Device available for evaluation: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in japan as follows: it was reported that the patient underwent a spinal fusion for a multiple vertebral fracture on (b)(6) 2019.After the surgery, the rod was broken between l4 and l5.Additionally, the plf-fusion bone was fractured.A revision surgery was performed to prevent paralysis in the future.The revision surgery was performed in sequence as follows: 1.The eight screws inserted in t10 to l1 were removed.The screws inserted in l2 to s2 were not removed.2.Four top notch connectors (product code: 179771555) were used to fix four rods (product code: 179762480).3.Two cross connectors were used to connect the two additional rods.The revision surgery was completed successfully with no delay.No further information is available.This report involves one viper 2 system rod, straight 480mm.This is report 2 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A manufacturing record evaluation was performed for the finished device product code: 186789480.Lot number: tbwrb.It was electronically reviewed and no non-conformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 15/08/2018.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device from attachments visual analysis of the photo revealed that the viper2 straight rod-480mm had broken in two pieces.X-ray evidence provided does not show clearly the breakage of the rod.Nevertheless, photos of the devices can confirm the complaint condition, several sections of the fracture surface appear to be smoothed out, most likely due to constant friction between the fragments before removal.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the viper2 straight rod-480mm.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that viper2 straight rod-480mm was broken.The broken fragment was returned for examination.The rod was also found severely bent.No other issues were found.A dimensional inspection for the viper2 straight rod-480mm was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the viper2 straight rod-480mm would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed? yes, reviewed.Dimensional inspection: n/a device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|