| Model Number |
A-SERIES LIGHTING SYSTEM |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported the light handle of their harmonyair a-series lighting system is detaching and falling during use.Injuries or procedure delays were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A steris field service technician arrived onsite to inspect the lighting system and found the light handle was not properly locking into position when installed on the lighthead resulting in the reported event.The lighting system has been removed from service pending repairs.A follow-up mdr will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The user facility reported the light handle of their harmonyair a-series lighting system is detaching and falling during use.No injuries or procedure delays were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation determined that the light handle locking mechanism was not fully engaged resulting in the reported event.The technician replaced the light handle, tested the unit, confirmed it to be operation according to specifications and returned it to service.A 3-year complaint review indicates this to be an isolated event.No additional issues have been reported.
|
| |
|