| Model Number |
IAS8-120LP |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
11/28/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is not expected to be returned for evaluation and review.However, the complaint investigation is not complete at this time.A supplemental and final report will be filed following the completion of the complaint investigation.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
The sales representative reported on behalf of the customer that the ias8-120lp, airseal 8/120mm port, was used during a gyn procedure on (b)(6) 2022 when it was reported, ¿airseal 8 mm trocar sound cap flaps broke off and fell in the patient.The surgeon was unable to retrieve all of the flaps and 3 of the 4 flaps were left in the patient¿.The procedure was completed without an alternative device and there was no report of a delay to the case.Further assessment questioning found that no imaging was done to help locate the 3 missing device caps and the patients status was, ¿okay¿.This was reported for a life threatening injury, but there was no report of medical intervention or prolonged hospitalization for the patient.This report is being raised on the basis of injury due to fragmentation left in the body.
|
| |
|
Additional Manufacturer Narrative
|
|
The device will not be returned, and no photographic evidence was provided.Therefore, the reported event cannot be verified.A device history record review found no abnormalities that would contribute to this reported event.A two-year lot history review shows this is the only complaint for this lot number and failure mode.A two-year review of complaint history revealed there has been a total of 27 reports, regarding 31 devices, for this device family and failure mode.(b)(6) per the instructions for use, the user is advised the following: the user is advised that if using the optional sound cap (8 mm, 12 mm): inspect sound cap blue foam and blue seal prior to use.Use caution when inserting a sharp or large device through the cannula.Inspect the sound cap after use for physical damage of any kind.Use extreme caution during airseal access port insertion.Improper use of this product can result in life-threatening injury to internal organs and vessels.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
The sales representative reported on behalf of the customer that the ias8-120lp, airseal 8/120mm port, was used during a gyn procedure on (b)(6) 2022 when it was reported, ¿airseal 8 mm trocar sound cap flaps broke off and fell in the patient.The surgeon was unable to retrieve all of the flaps and 3 of the 4 flaps were left in the patient.¿.The procedure was completed without an alternative device and there was no report of a delay to the case.Further assessment questioning found that no imaging was done to help locate the 3 missing device caps and the patients status was, ¿okay¿.This was reported for a life threatening injury, but there was no report of medical intervention or prolonged hospitalization for the patient.This report is being raised on the basis of injury due to fragmentation left in the body.
|
| |
|