INTEGRA LIFESCIENCES SWITZERLAND SAR CRTS SM IL VLV W CATH ACC; CERTAS PLUS SMALL & RA Back to Search Results
Model Number 82-8811PL
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/12/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A facility reported a certas valve (id 828811pl) was implanted on (b)(6) 2022 with a set pressure of 7.On (b)(6) 2022, valve was found to be incorrectly set to 3.The patient had symptoms of over-drainage.A 3t mri was performed on (b)(6) 2022 prior on finding the incorrect setting.The valve was removed on (b)(6) 2022.
 
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
The certas valve (id 828811pl) was returned for evaluation.Dhr - lot 4499922, conformed to the specifications when released to stock failure analysis - the position of the cam when valve was received was at setting 7.The valve was visually inspected; no defect was noted.The valve was hydrated.The valve passed the test for programming, occlusion, leak, reflux, siphon guard and pressure.No root cause could be determined as the technician could not confirm any problem with the valve at the time of investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRTS SM IL VLV W CATH ACC
Common Device Name
CERTAS PLUS SMALL & RA
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key16064203
Report Number3013886523-2022-00586
Device Sequence Number20509567
Product Code JXG
UDI-Device Identifier10886704081906
UDI-Public(01)10886704081906(17)250131(10)4499922
Combination Product (Y/N)N
PMA/510(K) Number
K182265
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Physician Assistant
Type of Report Initial,Followup
Report Date (Section B) 04/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model Number82-8811PL
Device Catalogue Number828811PL
Device Lot Number4499932
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/14/2022
Supplement Date Received by Manufacturer04/12/2023
Initial Report FDA Received Date12/28/2022
Supplement Report FDA Received Date04/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/26/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-