• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SENSOR; STYLET FOR CATHETER, GASTRO-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SENSOR; STYLET FOR CATHETER, GASTRO-UROLOGY Back to Search Results
Model Number M0066703051
Medical Device Problem Codes Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported to boston scientific corporation that a sensor wire guidewire was used on (b)(6) 2022.During the procedure, the blue part of the device had come away and detached from the core wire by approximately 10 cm.The procedure was completed with another of the same sensor wire guidewire there were no patient complications reported as a result of this event.
 
Event or Problem Description
It was reported to boston scientific corporation that a sensor wire guidewire was used on (b)(6) 2022.During the procedure, the blue part of the device had come away and detached from the core wire by approximately 10 cm.The procedure was completed with another of the same sensor wire guidewire there were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6 (device codes): device code a0401 captures the reportable event of corewire break.Block h10: the returned sensor wire guidewire was analyzed.Upon visual assessment, it was observed that the sleeve detached.The sleeve was not return.The core wire break was not confirmed because there was no evidence of this damage.Based on the condition of the returned device, engineers determined that excess force was applied in removing the wire which probably caused the detachment of the sleeve, it is most likely that as part of the device manipulation during the procedure an excess of force was applied to the device such as during the guidewire insertion through another device or the interaction with the scope, causing the sleeve detached.Boston scientific has determined the most probable cause of this complaint is adverse event related to procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR
Common Device Name
STYLET FOR CATHETER, GASTRO-UROLOGY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key16064405
Report Number3005099803-2022-07774
Device Sequence Number19335248
Product Code EZB
UDI-Device Identifier08714729302650
UDI-Public08714729302650
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/07/2025
Device Model NumberM0066703051
Device Catalogue Number670-305
Device Lot Number0028832848
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/08/2022
Supplement Date Received by Manufacturer01/24/2023
Initial Report FDA Received Date12/28/2022
Supplement Report FDA Received Date02/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-