STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5_12.6
Medical Device Problem Codes Misfocusing (1401); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Claim # (b)(4).
 
Event or Problem Description
The reporter indicated the surgeon implanted a 12.6mm vticm5_12.6 implantable collamer lens, -09.50/+3.0/048 (sphere/cylinder/axis) in the patients left eye (os) on (b)(6) 2022.The patient experienced a refractive surprise and the lens remained implanted.Additional information has been requested but none has been forthcoming.
 
Additional Manufacturer Narrative
Claim#: (b)(4).
 
Additional Manufacturer Narrative
Additional information: h6-health effect- impact code: "2199" should be removed and code "4629" should be added.H6- medical device problem code: "2993" should be added.B5- the reporter indicated the surgeon implanted a 12.6mm vticm5_ 12.6 implantable collamer lens of diopter -09.50/+3.0/048 (sphere/cylinder/axis) in the patient's left eye (os) on (b)(6) 2022.The patient experienced a refractive surprise.On (b)(6) 2023 the lens was exchanged for the same model, length and the problem was resolved.Status of the eye: "no signs".The cause of the event is unknown.Claim# (b)(4).
 
Additional Manufacturer Narrative
Device evaluation: h3 - type of investigation code: 10- device evaluation: the lens was returned dry in a microcentrifuge vial with residue on product.Visual inspection found haptic bent.Dimensional inspection found the lens to be within specifications.Functional inspection found the returned lens did not meet original values measured at the time of manufacturing.Additonal information: h6- type of investigation code: 3331- device history record (dhr) review: based on the results of the investigation, all released devices from the associated work order(s), including the suspected device, have been manufactured within established process parameters; and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene enriquez
1911 walker avenue
monrovia, CA 91016
MDR Report Key16064758
Report Number2023826-2022-04682
Device Sequence Number12438772
Product Code QCB
UDI-Device Identifier00840311308934
UDI-Public00840311308934
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Model NumberVTICM5_12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2022
Supplement Date Received by Manufacturer01/23/2023
06/27/2023
10/19/2023
Initial Report FDA Received Date12/28/2022
Supplement Report FDA Received Date01/23/2023
06/28/2023
10/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age27 YR
Patient SexMale
-
-