Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional pro-code: ktt.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Investigation summary: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that on (b)(6) 2022, during intra-op plate was placed for a proximal humerus fracture.The most top anterior screw hole was drilled with the locking drill guide, but the locking screw stripped when it engaged with the plate.A different plate was placed without issue.Surgery was completed successfully.There was no surgial delay.Procedure was completed successfully.There was no fragments generated.There was no patient consequences.This complaint involves two (2) devices.This report is for one (1) 3.5 lckng screw slf-tpng w/sd(tm) rec 28.This is report 2 of 2 for complaint (b)(4).
|