Model Number BSM-3532A
Medical Device Problem Codes Mechanical Problem (1384); Detachment of Device or Device Component (2907); Device Fell (4014)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported that the bedside monitor (bsm) fell from the wall mount it was attached to.The mount was not correctly mounted into the wall studs, and the device fell from about 5 feet up.No harm or injury was reported.They would like to send the unit in to fix the physical damage.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.Additional device information: concomitant medical device: the following device was used in conjunction with the bsm and is the device that experienced the failure: wall support bracket: model #: the customer could not provide this information.Serial #: the customer could not provide this information.
 
Event or Problem Description
The customer reported that the bedside monitor (bsm) fell from the wall mount it was attached to.The mount was not correctly mounted into the wall studs, and the device fell from about 5 feet up.No harm or injury was reported.
 
Event or Problem Description
The customer reported that the bedside monitor (bsm) fell from the wall mount it was attached to.The mount was not correctly mounted into the wall studs, and the device fell from about 5 feet up.No harm or injury was reported.
 
Additional Manufacturer Narrative
Details of complaint: the customer reported that the bedside monitor (bsm) fell from the wall mount it was attached to.The mount was not correctly mounted into the wall studs, and the device fell from about 5 feet up.No patient harm or injury was reported.Investigation summary: the customer stated that their maintenance personnel installed the mount and that the mount was not correctly mounted into the studs of the wall.No injuries were reported.Based on the available information, the most probable cause of the issue is user error - incorrect installation.Should the mount have been properly installed, the issue may have been prevented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BSM-3532A
Common Device Name
VITAL SIGNS MONITOR
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key16065194
Report Number8030229-2022-03265
Device Sequence Number2034031
Product Code MHX
UDI-Device Identifier04931921113660
UDI-Public04931921113660
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K080342
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBSM-3532A
Device Catalogue NumberBSM-3532A
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2022
Supplement Date Received by Manufacturer03/09/2023
Initial Report FDA Received Date12/28/2022
Supplement Report FDA Received Date04/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/14/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
WALL SUPPORT BRACKET; WALL SUPPORT BRACKET
Patient SexUnknown
-
-