| Model Number |
VTICM512.1 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 12.6mm vticm512.6 implantable collamer lens of -11.0/2.0/074 (sphere/cylinder/axis) was implanted into the patient's eye.On (b)(6) 2022, the lens was exchanged due to the lens being too small and lens rotation.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data b5: the reporter indicated that a 12.1mm vticm512.1 implantable collamer lens of -11.0/1.5/086 (sphere/cylinder/axis) was implanted into the patient's eye on (b)(6) 2022.On (b)(6) 2022, the lens was exchanged for a longer length lens due to the lens being too small and lens rotation.Reportedly, "iop was good." claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B5: corrected to: the reporter indicated that a 12.1mm vticm512.1 implantable collamer lens of -11.0/1.5/086 (sphere/cylinder/axis) was implanted into the patient's eye on (b)(6) 2022.On (b)(6) 2022, the lens was explanted and replaced for a longer length lens due to the lens being too small and lens rotation.Reportedly, "iop was good." h6 - health effect - impact code: 4627 should be added to initial mdr.H6 - health effect - impact code: 4629 should be removed from the initial mdr.Claim#: (b)(4).
|
| |
|