• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC NAVISTAR¿ ELECTROPHYSIOLOGY CATHETER; CATHETER, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC NAVISTAR¿ ELECTROPHYSIOLOGY CATHETER; CATHETER, ELECTRODE RECORDING Back to Search Results
Model Number NS7TCDL174HS
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Heart Block (4444)
Date of Event 12/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device investigation details: information received indicates that the device is not available for return, therefore no product investigation can be performed, and the customer complaint cannot be confirmed.A manufacturing record evaluation was performed for the finished device number lot 30843317m and no internal actions related to the complaint was found during the review.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.(b)(6).If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Event or Problem Description
It was reported that a male patient underwent a paroxysmal supraventricular tachycardia (psvt) ablation procedure with a navistar¿ electrophysiology catheter.The patient suffered heart block av requiring temporary pacemaker and prolonged hospitalization.It was reported in a psvt procedure, av block was developed during ablation of the slow pathway region.Issue occurred within 30 minutes after the catheter use started.After the procedure, a temporary pacemaker was inserted, and the patient was returned to the room.The patient developed complete atrioventricular block (junctional rhythm) during the procedure.The physician's opinions on the relationship between the event and the product was that the cause was the procedure and physician¿s own judgment error.During the ablation, the physician paid attention to various points, and the judgment was delayed.Ablation of the slow pathway region was successful but resulted in avb.Additional information received indicated the adverse event was discovered during use of biosense webster products.The patient required extended hospitalization because of the adverse event.The outcome from the adverse event was reported as fully recovered and the patient was discharged from the hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVISTAR¿ ELECTROPHYSIOLOGY CATHETER
Common Device Name
CATHETER, ELECTRODE RECORDING
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key16065809
Report Number2029046-2022-03251
Device Sequence Number2009882
Product Code DRF
UDI-Device Identifier10846835000597
UDI-Public10846835000597
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P990025
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/12/2025
Device Model NumberNS7TCDL174HS
Device Catalogue NumberNS7TCDL174HS
Device Lot Number30843317M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2022
Initial Report FDA Received Date12/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
8.5F SHEATH WITH CURVE VIZ SMC; DEF 5FR,CC,10P,P2,2MM LE,5-15; OCTA,LNG,48P,3-3-3-3-3,D-CURVE; SMARTABLATE GENERATOR KIT-WW
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age65 YR
Patient SexMale
-
-