Model Number 0602833
Device Problems Fracture (1260); Material Separation (1562); Dent in Material (2526)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/05/2022
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 12/2023).
 
Event Description
It was reported that two years one month post a port placement, the catheter was allegedly found to be broken.It was further reported that the distal catheter segment was removed.There was no reported patient injury.
 
Event Description
It was reported that approximately two years and one month post port placement procedure, the catheter was allegedly found to be broken.It was further reported that the distal catheter segment and port body were removed.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one groshong catheter was returned for evaluation.Functional, gross visual, tactile and microscopic visual evaluations were performed.The investigation is confirmed for the reported catheter fracture, material separation and the identified dent in material issues, as a complete circumferential break was noted on the proximal end of the catheter which was elliptical in shape.Furthermore, a partial circumferential break was noted on the catheter approximately 1.3cm from the proximal end and a split was noted on the catheter approximately 2.2cm from the proximal end.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 12/2023).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.R.I. IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, KIT, 8F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key16065827
MDR Text Key308257391
Report Number3006260740-2022-05889
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741025723
UDI-Public(01)00801741025723
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K873213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0602833
Device Catalogue Number0602833
Device Lot NumberREEN1471
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/07/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-