|
C.R. BARD, INC. (BASD) -3006260740 M.R.I. IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, KIT, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
| Model Number |
0602833 |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to a port placement procedure, foreign material like hair was allegedly found in the kit.There was no patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, photos were provided for review.The investigation of the reported event is currently underway.Expiry date: 09/2025.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned for evaluation.However, four electronic photos were provided for review.The photo shows one sample tray containing one introducer needle, one 12cc syringe and one tunneler.Upon zooming the picture, a hair strand was noted on the sample tray.Manufacturing site evaluation of the sample found that inside the packaging near the distant tip of the tunneler there was a black hair that was making contact with this component.Therefore the reported contamination issue can be confirmed.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.D4 (expiry date: 09/2025).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to a port placement procedure, foreign material like hair was allegedly found in the kit.There was no patient contact.
|
| |
|
Search Alerts/Recalls
|
|
|