On (b)(6)2022 getinge became aware of an issue with one of surgical lights - powerled ii.It was stated screw connection of the central axis in the ceiling tube is defective.All internal threads in the central axle are damaged and none of the 6 screws hold position causing uneven connection.There was no injury reported, however, we decided to report the issue in abundance of caution as this unstable connection could lead to fall of the device and as a result of that, could lead to serious injury.
|
|
The correction of h4 device manufacture date field deems required.This is based on the internal evaluation.Previous h4 device manufacture date: 25th september 2020.Corrected h4 device manufacture date: 28th september 2020.The correction of h3a device evaluated by manufacturer, h3b device not eval provide code, h3c if other provide code - explain fields deems required.This is based on the internal evaluation.Previous h3a device evaluated by manufacturer: no.Corrected h3a device evaluated by manufacturer: yes.Previous h3b device not eval provide code: other.Corrected h3b device not eval provide code: n/a.Previous h3c if other provide code - explain: device not returned to manufacturer.Corrected h3c if other provide code - explain: n/a.The correction of d4 catalog # and unique identifier (udi) # deems required.This is based on the internal evaluation.Previous d4 catalog # ardpwt209142a.Corrected d4 catalog # blank.Previous d4 unique identifier (udi) #: n/a.Corrected d4 unique identifier (udi) #: (b)(4).The correction of h6 medical device ¿ problem code deems required.This is based on the internal evaluation.Previous h6 medical device ¿ problem code: mechanical problem|unintended.Movement|unstable|1667.Corrected h6 medical device ¿ problem code: mechanical problem|detachment of device or device component||2907, material integrity problem|degraded|peeled/delaminated|1454.The correction of d1 brand name deems required.This is based on the internal evaluation.Previous d1 brand name powerled ii 500.Corrected d1 brand name maquet powerled ii.Getinge became aware of an issue with one of surgical lights ¿ powerled ii.It was stated screw connection of the central axis in the ceiling tube is defective.All internal threads in the central axle are damaged and none of the 6 screws hold position causing uneven connection.Additionally, based on photographic evidence, a label chipping and paint peeling was observed on suspension arm.There was no injury reported, however, we decided to report the issue in abundance of caution as this unstable connection could lead to fall of the device and as a result of that, could lead to serious injury.Based on the information collected, it was established that when the event occurred, the surgical light did not meet its specification and in this way the device contributed to event.The claimed device was not being used for patient treatment or diagnosis when the event took place.It was found during maintenance of the device.The cause of the tapping defect is due to the use of an incorrect tool pairing when machining axis ard567802131.By following the instructions in the maquet powerled ii ni 01814 en 04 installation manual, this fault can be detected during installation.Below are the instructions and warning included in the installation manual: ¿using a 5 mm allen key, tighten the six screws, starting with the two screws on the same side.Use a torque wrench to finish tightening to a torque of 16 n.M (11.80 lb.Ft)¿.¿danger! risk of death: the patient or user may be in danger if the tightening torques are not observed (e.G.: risk of falling, electrical hazard, etc.) it is essential to apply the tightening torque(s) specified in this section¿.¿modifications to the product: getinge may not be held liable for any modifications made to the product that are not expressly approved by getinge¿.¿warning! risk of infection: lightweight parts from the device may fall onto the surgical site.Check that all fastenings, covers, cover plates and bumpers on the device are properly in place¿.The large mark on the suspension arm that damaged the paint and the label was probably caused by rubbing during mishandling.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
|