MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754CML; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754CML; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-754CML
Device Problems Premature Discharge of Battery (1057); Loss of Data (2903)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2022
Event Type  malfunction  
Event Description
Information received by medtronic indicated that the customer reported that there was an "a21 alarm - unexpected restart and a cold reset were performed" and an "a17 alarm - external ram crc error" on the insulin pump.No harm requiring medical intervention was reported.Troubleshooting was performed and the customer was able to clear the alarm and rewind the insulin pump but the error table indicated that the insulin pump should be replaced.The customer will discontinue using the device and the insulin pump will be returned for analysis.
 
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754CML
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*   00777-3869
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
*   00777-3869
7635140379
MDR Report Key16067002
MDR Text Key308188461
Report Number2032227-2022-390983
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-754CML
Device Catalogue NumberMMT-754CML
Device Lot NumberA7754CMLJ
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/21/2022
Initial Date FDA Received12/29/2022
Date Device Manufactured09/30/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-