| Model Number |
101/860/085CZ |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that during the pre-use check, leakage of air from the cuff was observed.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: h6: event problem and evaluation codes: updated h10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.One not used sample was received in plastic bag without its original packaging.Under visual inspection the sample appeared to be in good condition.During manufacturing process the devices are inflation tested, which includes inflating each device cuff and leaving for a 12 hour period.Reductions in pressure over this time are considered a failure and the device would be rejected.The inflation test was repeated on the received sample.It was confirmed that after 12 hours the cuff was still fully inflated.Based on results of testing the reported failure was not confirmed.The cause of the reported problem could not be determined.A dhr (device history review) was not performed because no fault was found.
|
| |
|