| Model Number |
M655G |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Component code: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: when was the dregs of suture products fall (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? during use on the patient the following information was requested, but unavailable: was there any adverse consequence associated with the patient? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on (b)(6) 2022 and suture was used.In the process of use, it is found that the dregs of suture products fall.Changed another one to complete the surgery.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).H6 component code: g07002 - no device problem found.H 6.Investigation findings: c22 ¿ photo investigation.H3 investigation summary: the product was returned for evaluation.Visual inspection evaluation was conducted on the returned device.Visual analysis of the sample determined that one opened sample that pertains to product code m655g was returned for analysis.The product code is multistrand and each packet contains four strands single armed.Upon initial inspection of the sample no foreign matter, pieces of debris, or defects were observed along the strands received.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.Photo investigation summary = > this is an analysis for a photo submitted for evaluation.During the visual analysis, the following was observed: the photo shows an unknown material on a surface in a surgery, the suture is not present in the photograph.The image is not clear to determine the failure mode or the reported condition.Based on the photo review, the event describe is not confirmed, however no conclusion or root cause could be determined.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected information: d 9.Device available for evaluation?, h 3.Device evaluated by manufacturer?.
|
| |
|
Search Alerts/Recalls
|
|
|