Model Number LF1937 |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problems
Unspecified Tissue Injury (4559); Insufficient Information (4580)
|
Event Date 12/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during laparoscopic pancreaticoduodenectomy, the jaw of the device was stuck after being closed and could not be opened in the middle of the entire surgery, tissue was eventually removed with other instrument, the knife blade was extended but not protruded beyond the edge of the jaws.There was no patient injury.
|
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The involved device was not returned.Visual inspection noted no conditions.Functional testing found the device's jaw opening and closing mechanism was functioning properly.The weld integrity of the device was inspected and was found to be within specification.The device was activated multiple times on a saline soaked gauze pad with satisfactory results.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The device was activated with the rotation knob in various positions to detect any problematic activation issues.No electrical or wiring issues were found.It was reported that the jaws of the device were difficult to open.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition: the unit had a knife failure.Functionally, the knife cut of the device was tested on a silicone test strip with unacceptable results.The knife-edge was found to be damaged during high mag inspection, and the device failed the cut test.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: avoid grasping objects, such as staples, clips, or encapsulated sutures in the jaws of the instrument.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|