• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. CLAREON VIVITY EXTENDED VISION UV ABSORBING IOL; EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. CLAREON VIVITY EXTENDED VISION UV ABSORBING IOL; EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS Back to Search Results
Model Number CCWET0
Medical Device Problem Codes Failure to Advance (2524); Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A registered nurse reported that during a cataract extraction with an intraocular lens (iol) implant procedure, lens became stuck in the cartridge and the insertion device scuffed the lens.There was patient contact observed.The surgery was completed the same day with another lens.Additional information was received stating no harm was done to the patient.
 
Additional Manufacturer Narrative
Additional information was provided in d.9, h.3, h.6 and h.10.The product was returned for analysis and damage was observed to the intraocular lens.No cartridge was returned.Intraocular lens returned pressed against one post of the lens case base resulting in damage to the intraocular lens.A significant amount of solution is dried on both surfaces of the optic and haptics.One haptic is adhered to the optic with solution, the other haptic is bent/deformed.The optic is cracked/fractured and scratched/marked-rejectable.Unable to conduct dimensional testing due to condition of returned sample.No cartridge received.The product investigation could not identify the root cause for the reported complaint "lens became stuck in the cartridge" as only the intraocular lens was returned for evaluation.Due to this, we are unable to complete a thorough investigation to determine a root cause.Investigation has been completed based on current information.No further action is warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON VIVITY EXTENDED VISION UV ABSORBING IOL
Common Device Name
EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16069124
Report Number9612169-2022-00738
Device Sequence Number20416482
Product Code POE
UDI-Device Identifier00380652405850
UDI-Public00380652405850
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/21/2025
Device Model NumberCCWET0
Device Catalogue NumberCCWET0.240
Device Lot Number25289876
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/08/2022
Supplement Date Received by Manufacturer03/24/2023
Initial Report FDA Received Date12/29/2022
Supplement Report FDA Received Date04/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CLAREON MONARCH IV, INJECTOR; MONARCH III, CARTRIDGE D; UNSPECIFIED LOADING FORCEPS; VISCOAT
Patient SexUnknown
-
-