• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US PENETRATING GRASPER 45° RIGHT; SUTURE/NEEDLE PASSER, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US PENETRATING GRASPER 45° RIGHT; SUTURE/NEEDLE PASSER, REUSABLE Back to Search Results
Model Number 214603
Medical Device Problem Codes Difficult to Open or Close (2921); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The expiration date is currently unavailable.Reporter is a j&j employee.Investigation summary: the product was returned to mitek for evaluation.Mitek then conducted visual inspection and functional test of device received.Visually, the device appears worn indicating device was heavily used; there were signs of wear on the device and the laser markings were slightly faded.Also, the jaw was loose, they were not functioning smoothly.It was observed that when the trigger of the device was pulled, the jaws would not actuate open/close consistently.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.Based on the visual results, this complaint can be confirmed.The lot number provided indicates this device is 13 years old and has seen considerable use in the field which is consistent with its appearance.This type of failure has been attributed to wear and tear as well as rough use, since the repeated excess force applied when a large amount of tissue is being grasped which cause a potential to break; also, the age of the device causing damage in the device.Other than this possibility, cannot be discerned at this point.As per ifu, inspect for damage before using.End of useful instrument life is generally determined by wear or damage from handling or surgical use.Inspect instruments between uses to verify proper functioning.Between uses, lubricate moving parts with a water-soluble lubricant in accordance with the manufacturer¿s instructions.At this point in time, no corrective action is required, and no further action is warranted.However, in depuy synthes mitek, additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
 
Event or Problem Description
It was reported by the affiliate in colombia that the device had an unspecified malfunction.During in-house engineering evaluation, it was determined that the penetrating grasper 45° right device appeared worn indicating device was heavily used, had signs of wear with the laser markings slightly faded.It was further determined that the jaw was loose and was not functioning smoothly; and when the trigger of the device was pulled, the jaws would not actuate open/close consistently.There was no procedure nor patient involvement reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENETRATING GRASPER 45° RIGHT
Common Device Name
SUTURE/NEEDLE PASSER, REUSABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key16069292
Report Number1221934-2022-04312
Device Sequence Number17707769
Product Code NBH
UDI-Device Identifier10886705003860
UDI-Public10886705003860
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number214603
Device Catalogue Number214603
Device Lot Number09G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/28/2022
Initial Report FDA Received Date12/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/04/2009
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-