Model Number GIF-XTQ160
Medical Device Problem Code Microbial Contamination of Device (2303)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Attempts to retrieve additional information from the customer are in progress.This event is under investigation.A supplemental report will be submitted upon receiving additional information.
 
Event or Problem Description
The customer reported the contamination levels of the evis exera gastrointestinal videoscope were too high.The issue occurred during reprocessing.There was no reported patient harm or impact due to this event.
 
Additional Manufacturer Narrative
This report is being supplemented to provide results of microbial testing and evaluation.After the device was returned to olympus, it was sent out for additional testing.The microbiological analysis report indicated the channels of the scope were cultured and the results obtained comply with the target level for an endoscope subjected to high level disinfection and rinsed with sterile water.During device evaluation at olympus, it was found the scope cover plate was detached, the suction cylinder was discolored, switch #1 was cut, the insulation resistance did not meet the standard value due to a crack on the scope cover, the bending angle in the up direction did not meet the standard value due to wear of the angle wire, the scope cover was deformed, the bending section cover adhesive was dirty, and the coating of the connecting tube was peeled.This event is under investigation.A supplemental report will be submitted upon receiving additional information.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation and because the subject device was not returned and the user culture results not shared, the reported event could not be confirmed and a root cause could not be determined.Growth of microorganisms were reported through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.The following is included in the device ifu: "an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them.".Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA GASTROINTESTINAL VIDEOSCOPE
Common Device Name
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16070108
Report Number9610595-2022-06271
Device Sequence Number6234529
Product Code FDS
UDI-Device Identifier04953170339837
UDI-Public04953170339837
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K011151
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGIF-XTQ160
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/04/2022
Supplement Date Received by Manufacturer02/07/2023
04/24/2023
Initial Report FDA Received Date12/29/2022
Supplement Report FDA Received Date03/08/2023
05/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/11/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-