Model Number 6371
Device Problems Fracture (1260); High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/12/2022
Event Type  Injury  
Event Description
It was reported during ipg replacement, the lead extension appeared fractured as the physician struggled to advance the extension into the header.Intra-op diagnostics measured high impedances.Multiple attempts were made to clear the impedances using the lead and extension insertion tool.The extension was inserted into the port measuring high impedances on one contact.Effective therapy was restored post-operatively.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8CH INFINITY DBS FLEX EXTN KIT, 50CM, B
Type of Device
DBS EXTENSION
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key16070908
MDR Text Key306361639
Report Number1627487-2022-07145
Device Sequence Number1
Product Code MHY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/09/2020
Device Model Number6371
Device Catalogue Number6371
Device Lot Number6509939
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/12/2022
Initial Date FDA Received12/29/2022
Supplement Dates Manufacturer Received03/03/2023
Supplement Dates FDA Received03/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/10/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexMale
Patient Weight84 KG
-
-