Model Number N/A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of regs3161 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported by the customer that the microintroducer sheath failed to insert passed the skin and folded back or began to split upon attempt to insert into the vein.No other information was provided.
 
Event or Problem Description
It was reported by the customer that the microintroducer sheath failed to insert passed the skin and folded back or began to split upon attempt to insert into the vein.No other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Based on a review of available complaint information and an evaluation of the available sample evidence, the following was concluded: the complaint of the microintroducer failing to insert passed the skin is confirmed and was determined to be due to insertion techniques.One 3.5 fr microintroducer was returned for evaluation.An initial visual observation showed blood use residues throughout the returned device and visual damage to the distal end of the gray microintroducer sheath.No obvious bends were noted along the microintroducer.A microscopic observation revealed deformation to the distal end of the gray sheath of the microintroducer, and a gradual taper was visible from manufacturing.The complaint of difficulty inserting a microintroducer is confirmed.Possible causes may include insertion procedure of the device or insertion angles of the device, as it may catch on the patient¿s skin or vessel walls.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOT 3FR SL PROV ML
Common Device Name
MIDLINE CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kayla olsen
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key16071016
Report Number3006260740-2022-05904
Device Sequence Number14999098
Product Code PND
UDI-Device Identifier00801741185076
UDI-Public(01)00801741185076
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K153393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberSP4153108DB
Device Lot NumberREGS3161
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/12/2022
Supplement Date Received by Manufacturer04/03/2023
Initial Report FDA Received Date12/29/2022
Supplement Report FDA Received Date04/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-