• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AUTOTOME RX 49; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AUTOTOME RX 49; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00545150
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Imdrf device code a0401 captures the reportable event of cutting wire broken.
 
Event or Problem Description
It was reported to boston scientific corporation that an autotome rx 49 was used in the duodenal papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, "the procedure could not be performed due to the cutting wire was fractured." the procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.No further information has been obtained despite good faith efforts.The reported event may suggest that the cutting wire broke.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Block h10: the returned autotome rx 49 was analyzed, and a visual evaluation noted that the wire/anchor got dislodged or dislocated from its distal pierced hole and the cutting wire was bent/kinked.The anchor was also broken, and the broken section was not returned.The device was observed under magnification, and both the distal and proximal pierced holes were torn, and the anchor was broken.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the anchor was broken, and the cutting wire/anchor dislodged or dislocated from its distal pierce hole, the cutting wire was bent/kinked, and the working length was torn at the distal and the proximal pierced holes.Based on the condition of the device, the problem found could have been generated when the cutting wire was submitted to tension during the handle actuation.Also, bowing the device without being completely out of the scope can lead to tear it, displacing the cutting wire anchor from its position.Additionally, it was found that the working length was torn at the proximal pierce hole and the cutting wire was kinked.Once the anchor got dislocated from the distal pierced hole during the procedure, the manipulation with additional tools/devices or the scope could lead to kinking the cutting wire, tearing the proximal pierced hole, consequently, breaking the anchor.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation that an autotome rx 49 was used in the duodenal papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, "the procedure could not be performed due to the cutting wire was fractured." the procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.No further information has been obtained despite good faith efforts.The reported event may suggest that the cutting wire broke.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOTOME RX 49
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key16073325
Report Number3005099803-2022-07854
Device Sequence Number12320646
Product Code KNS
UDI-Device Identifier08714729415138
UDI-Public08714729415138
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K013153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/02/2024
Device Model NumberM00545150
Device Catalogue Number4515
Device Lot Number0028497563
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/09/2022
Supplement Date Received by Manufacturer01/16/2023
Initial Report FDA Received Date12/29/2022
Supplement Report FDA Received Date02/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
Patient Weight70 KG
-
-