Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Block h10: the returned autotome rx 49 was analyzed, and a visual evaluation noted that the cutting wire was broken, and it was not returned.The device was observed under magnification, and the cutting wire was broken, and the proximal end of the wire was blackened, which indicates that electric current was applied.No other problems with the device were noted.The reported event of the cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was broken, and blackened.The cutting wire being blackened indicates that the device was energized.Based on the condition of the device, the problem found could have been generated if there was contact between the device and the scope during energization or if the device exceed the maximum of voltage during procedure as per precautions of the device states, also energizing the device prior to performing sphincterotomy can compromise the cutting wire's integrity, also it can cause a premature cutting wire fatigue, leading to break it.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|