Model Number WA4KL530 |
Device Problem
Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/29/2022 |
Event Type
malfunction
|
Event Description
|
The customer reported to olympus, the telescope "ultra", 5.4 mm, 30° had a crack on the lens.There were no reports of patient harm associated with this event.
|
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation.Upon inspection testing of the returned device, the user's request was confirmed.It was discovered, there were broken optical lenses, a bent outer tube, debris under the window, dents on the funnel and minor dents on the distal edge.The investigation is ongoing.If additional information becomes available, this report will be supplemented accordingly.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide results of the investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.There was no indication that the event was caused by a misuse or that the event was related to design of the device.Repair history was reviewed and no issues were found related to the reported event.Although a definitive root cause of the defects could not be identified, it was determined that the broken lens was likely caused by excessive force during use by the customer.If additional information becomes available at a later date, this report will be supplemented.Olympus will continue to monitor the field performance of this device.
|
|