Model Number 110-360
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/25/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation: rupture.
 
Event or Problem Description
Healthcare professional reported right side "breast deformity with the right implant appearing smaller than the left," baker grade ii capsular contracture of the right breast," and patient also has textured mammary implants.Healthcare professional later reported "any creases" and "hole, non specific" and "hole in radius edge." device has been explanted and replaced.
 
Event or Problem Description
Healthcare professional reported, right side "breast deformity.With the right implant appearing smaller than the left".Baker grade ii, capsular contracture of the right breast".And patient also, has textured mammary implants.Healthcare professional later reported, "any creases" and "hole, non specific" and "hole in radius edge".Device has been explanted and replaced.
 
Additional Manufacturer Narrative
Device evaluation: based on the device analysis grid, the assessments of the complaint are: capsular contracture: unable to confirm, as it is a medical event.Not related to the device.Material rupture: observed, broken striated on anterior side assessed, as surgical damage.Anxiety: unable to confirm, as it is a medical event.Not related to the device.Crease fold: observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 110 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key16075414
Report Number9617229-2022-32700
Device Sequence Number6092981
Product Code FTR
UDI-Device Identifier10888628002241
UDI-Public10888628002241
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/22/2017
Device Model Number110-360
Device Catalogue Number110-360
Device Lot Number2263081
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/20/2021
Supplement Date Received by Manufacturer02/22/2023
Initial Report FDA Received Date12/29/2022
Supplement Report FDA Received Date02/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/22/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age41 YR
Patient SexFemale
Patient Weight58 KG
Patient RaceWhite
-
-