| Model Number |
SRX-255 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported left-side capsular contracture baker grade iii/iv.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter email: (b)(6).A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii/iv.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: capsular contracture: unable to observe as it is a medical event not related to the device.Additional observations: observed an opening on anterior side assessed as surgical damage and four openings on anterior side assessed as surgical damage like.Observed non penetrating nick on the device.Observed creases on the device.No further actions are required as the device was implanted.
|
| |
|