MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH ELECTROSURGICAL GENERATOR "ESG-400"; HF-GENERATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH ELECTROSURGICAL GENERATOR "ESG-400"; HF-GENERATORS Back to Search Results
Model Number WB91051W
Device Problems Output Problem (3005); Activation Problem (4042)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2022
Event Type  malfunction  
Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that during an unspecified time the esg-400 hf-generator issued error code e433.The issue was resolved by the customer a few days after the event occurred.No further information was provided but there was no report about an adverse event or patient injury.
 
Manufacturer Narrative
The suspect medical device was not returned to the manufacturer for evaluation/investigation.Therefore, the investigation/evaluation was performed exclusively on the basis of the information provided by the customer.The reported error message e433 is triggered by the generator¿s safety system and can have different technical causes.Since the generator was not returned to any olympus organization, the cause for the occurrence of the error message could not be determined in this case and it is also unknown whether the reported error was a permanent or a temporary fault.However, in the case of critical errors, the safety system will not permit any further use of the generator until the error is rectified.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected serial number of the hf generator without showing any abnormalities.The case will be closed on olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTROSURGICAL GENERATOR "ESG-400"
Type of Device
HF-GENERATORS
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
rheinstrasse 8
teltow
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662
MDR Report Key16076431
MDR Text Key308467796
Report Number9610773-2022-00806
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K203682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWB91051W
Device Catalogue NumberWB91051W
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/19/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse