Model Number 23-07-50
Medical Device Problem Codes Over-Sensing (1438); Low Readings (2460)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There was no patient involvement.Sn and udi are unknown.Since sn is unknown, also manufacturing date is unknown.Livanova deutschland manufactures the bubble detector.The incident occurred in united states of america.Livanova is following up with the customer.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event or Problem Description
Livanova deutschland received a report of bubble sensor issues on s5 hearth lung machine during service there was no patient involvement.
 
Additional Manufacturer Narrative
H10: through follow-up communication livanova learned that the unit was already replaced, the alarm would be intermittent and it was an over-sensing issue.Therefore, the event has been re-assessed as not reportable since no malfunction occurred.In case of bubble sensor triggering false bubble alarms, the pump is stopped by the bubble alarm even if no air is present in the monitored line.In such situations, the alarm can always be cleared/overridden by user, once the line is verified to be free of air, and the perfusion can be easily restarted.Thus, it is unlikely that a false bubble alarm will result in serious injury.
 
Event or Problem Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BUBBLE SENSOR FOR 3/8 TUBING
Common Device Name
MONITOR AND/OR CONTROL, LEVEL SENSING, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key16076518
Report Number9611109-2022-00697
Device Sequence Number15009746
Product Code KRL
UDI-Device Identifier04033817900597
UDI-Public(01)04033817900597(11)171120
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K955152
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 12/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number23-07-50
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2022
Supplement Date Received by Manufacturer02/23/2023
Initial Report FDA Received Date12/30/2022
Supplement Report FDA Received Date03/22/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-