• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065751155
Device Problem Suction Failure (4039)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported aspiration failure occurred during ultrasound (us) phase of cataract surgery.The surgery was completed after replacing the product with another one.There was no patient harm.
 
Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.10.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The returned sample was visually inspected and no obvious defects were found.A calibrated console representing the current software version was used to test the sample.The ball in the drip chamber¿s check valve moved freely per specification.The sample could prime and tune with the ultrasonic handpiece successfully.The irrigation and aspiration pressure were measured at multiple set points and met specifications.No message code appeared on the screen.Fluid flowed from balance salt solution (bss) to the drain bag without any interfering.No leakage was detected from the pump elastomer or on the pump area of the fluidics module.Cleaning process was able to be performed after functional test was completed.The sample passed functionality.The root cause of the customer's complaint could not be established; the returned sample functioned per specifications.After investigation of this complaint, it has been determined that this sample functioned per specifications; therefore, no corrective action is required at this time.Based on our current tracking, there are no adverse trends for this reported complaint and lot.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PACK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16077176
MDR Text Key308454737
Report Number1644019-2022-01097
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2024
Device Catalogue Number8065751155
Device Lot Number14C7T7
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/28/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CONSTELLATION VISION SYSTEM
-
-