INSULET CORPORATION OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN Back to Search Results
Model Number 18320
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/11/2022
Event Type  malfunction  
Event Description
It was reported that the pink slide did not move forward when the pod was activated; this indicates that a needle mechanism failure occurred.
 
Manufacturer Narrative
According to the complainant the device will not be returned for investigation.We are unable to confirm or determined the needle mechanism failure reported.Lot release records were reviewed and the product lot met all acceptance criteria.Specifically, a pod is paired to a pdm and put through simulated use testing including communicating with the pdm, deployment, delivering fluid, occlusion detection, and freedom from hazard alarms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
michael spears
100 nagog park
acton, MA 01720
9786007000
MDR Report Key16077432
MDR Text Key306608081
Report Number3004464228-2022-24798
Device Sequence Number1
Product Code LZG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192659
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 12/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/19/2023
Device Model Number18320
Device Catalogue NumberBLE-I1-529
Device Lot NumberPD1K11192111
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/11/2022
Initial Date FDA Received12/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/19/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age37 YR
Patient SexMale
-
-