TELEFLEX MEDICAL PE SAHARA DRY SUCT/DRY SEAL LF 6/CS; BOTTLE, COLLECTION, VACUUM Back to Search Results
Model Number IPN028484
Medical Device Problem Codes Crack (1135); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device history record of batch number 74e2000906 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non-conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled and inspected according to our specifications.
 
Event or Problem Description
Reported issue: incident occurred on (b)(6) 2022 in the pneumo- immuno allergology department.A pleur-evac drain was inserted in a patient suffering from a pleural effusion.When unclamping the tubing, pleural fluid went into the device, the doctor found a leak, and observed a crack in the tubing at the level of the clamp.There were no consequences for the patient.This is an isolated event.
 
Additional Manufacturer Narrative
Qn#(b)(4).One unit of catalog number s-1100-08lf pe sahara dry suct/dry seal lf 6/cs lot 74e2000906 was received for analysis.Sample was not received in its original packaging, and it shows sign of usage.The tubing section where the ats connector (red connector) is was not received.No cracks were observed in the tubing where the blue connector is located.A tubing section with the red ats connector from regular production was connected to the sample received to perform a functional test.No leaks were observed.The device history record has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non-conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled and inspected according to our specifications.No corrective action can be established since no problem found on sample.Customer complaint cannot be confirmed since no problem found on sample.
 
Event or Problem Description
Reported issue: incident occurred on (b)(6) 2022 in the pneumo- immuno allergology department.A pleur-evac drain was inserted in a patient suffering from a pleural effusion.When unclamping the tubing, pleural fluid went into the device, the doctor found a leak, and observed a crack in the tubing at the level of the clamp.There were no consequences for the patient.This is an isolated event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PE SAHARA DRY SUCT/DRY SEAL LF 6/CS
Common Device Name
BOTTLE, COLLECTION, VACUUM
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
effie jefferson
3015 carrington mill blvd
morrisville 27560
9194332672
MDR Report Key16077738
Report Number3004365956-2022-00121
Device Sequence Number12335905
Product Code KDQ
UDI-Device Identifier34026704631668
UDI-Public34026704631668
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/15/2023
Device Model NumberIPN028484
Device Catalogue NumberS-1100-08LF
Device Lot Number74E2000906
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2022
Supplement Date Received by Manufacturer02/22/2023
Initial Report FDA Received Date12/30/2022
Supplement Report FDA Received Date03/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/11/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED; NONE REPORTED
Patient SexUnknown
-
-