Model Number 26818 |
Device Problem
Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/25/2022 |
Event Type
malfunction
|
Event Description
|
The customer reported the screen flickers and power drops out.No patient impact or consequences were reported.
|
|
Manufacturer Narrative
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow-up report will be submitted.Initial reporter zip code exceeded the maximum allowable characters, zip code is as follows: (b)(6).
|
|
Event Description
|
The customer reported the screen flickers and power drops out.No patient impact or consequences were reported.
|
|
Manufacturer Narrative
|
Additional manufacturing narrative: other, other text: the returned device was evaluated.The device remained powered on and the display was stable during testing.Contaminated components observed on the instrument board potentially resulted in the reported issue.Initial reporter zip code exceeded the maximum allowable characters, zip code is as follows: (b)(6).
|
|