| Model Number |
4410 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned device was evaluated.When co2 gas mixture was applied, the isa module measured outside the accuracy specification.After gas span calibration zeroing was performed and a co2 gas mixture was applied, the isa module measured within the accuracy specification.Incorrect gas span calibration zeroing resulted in measurements outside accuracy specification.S.
|
| |
|
Event or Problem Description
|
|
The customer reported "bad results / reading." no patient impact or consequences were reported.
|
| |
|