| Model Number |
D012 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Delayed Charge Time (2586); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited a charge time of over fifteen seconds.Due to this, the device exhibited premature battery depletion.This device was explanted and replaced.No further adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned defibrillator was analyzed and an engineering-level longevity prediction calculation was completed to assess the rate of battery depletion.Given the programmed parameters and other data stored within the memory of the device, the results of this calculation indicated that the actual rate of battery depletion fell within an acceptable range.Next, a series of diagnostic tests were conducted that verified the performance of pacing, sensing, defibrillation, and recording functions.Having met the engineering longevity prediction, functionally passed all returned product testing, and with no further information to indicate a product performance issue, we have concluded that this device experienced normal battery depletion.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited a charge time of over fifteen seconds.Due to this, the device exhibited premature battery depletion.This device was explanted and replaced.This device was returned to boston scientific for analysis.No further adverse patient effects were reported.
|
| |
|