| Model Number |
CD3365-40C |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information was requested but not received.A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The product was returned.Interrogation and visual inspection were normal.Analysis found that the device was received with battery voltage above elective replacement indicator (eri) level.The cause of infection could not be traced to the device.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference number: 2017865-2023-00069; related manufacturer reference number: 2017865-2023-00071.It was reported that the patient presented in the hospital with septic shock due to bacteremia.Upon examination, the right atrial (ra) lead was discovered to have vegetation.The entire implantable cardioverter defibrillator (icd) system was explanted due to infection.There were no patient consequences.
|
| |
|