| Model Number |
IPN922773 |
| Medical Device Problem Code |
Deflation Problem (1149)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reported issue: the cuff stayed in the red zone when inflated then it was unable to be inflated or deflated with syringe.The defect was detected just before insertion into the patient.The intervention was to grab another one and keep bagging.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The device history was reviewed and there was no abnormality found.Based on the complaint description of "will not deflate prior to use", reviewing of the dhr of the packaging and assembly lot, reviewing the nc list, it was concluded that the complaint root cause is undetermined / unknown as the assembly or packaging process for the affected lot manufacturing process did not have any record on the similar issue.Further investigation shall be performed once the complaint sample was returned to kulim.
|
| |
|
Event or Problem Description
|
|
Reported issue: the cuff stayed in the red zone when inflated then it was unable to be inflated or deflated with syringe.The defect was detected just before insertion into the patient.The intervention was to grab another one and keep bagging.
|
| |
|