• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL LMA UNIQUE (SILICONE CUFF) AND LMA UNIQUE (SIL; AIRWAY, OROPHARYNGEAL, ANESTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL LMA UNIQUE (SILICONE CUFF) AND LMA UNIQUE (SIL; AIRWAY, OROPHARYNGEAL, ANESTH Back to Search Results
Model Number IPN922773
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
Reported issue: the cuff stayed in the red zone when inflated then it was unable to be inflated or deflated with syringe.The defect was detected just before insertion into the patient.The intervention was to grab another one and keep bagging.
 
Additional Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Additional Manufacturer Narrative
Qn#(b)(4).The device history was reviewed and there was no abnormality found.Based on the complaint description of "will not deflate prior to use", reviewing of the dhr of the packaging and assembly lot, reviewing the nc list, it was concluded that the complaint root cause is undetermined / unknown as the assembly or packaging process for the affected lot manufacturing process did not have any record on the similar issue.Further investigation shall be performed once the complaint sample was returned to kulim.
 
Event or Problem Description
Reported issue: the cuff stayed in the red zone when inflated then it was unable to be inflated or deflated with syringe.The defect was detected just before insertion into the patient.The intervention was to grab another one and keep bagging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LMA UNIQUE (SILICONE CUFF) AND LMA UNIQUE (SIL
Common Device Name
AIRWAY, OROPHARYNGEAL, ANESTH
Manufacturer (Section D)
TELEFLEX MEDICAL
athlone
Manufacturer (Section G)
THE LARYNGEAL MASK COMPANY
6 battery road #07-02
singapore 04990 9
SN   049909
Manufacturer Contact
effie jefferson
3015 carrington mill blvd
morrisville 27560
9194332672
MDR Report Key16085635
Report Number9681900-2023-00001
Device Sequence Number15136789
Product Code CAE
UDI-Device Identifier15060112318198
UDI-Public15060112318198
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN922773
Device Catalogue Number105200-000050
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/15/2022
Supplement Date Received by Manufacturer02/03/2023
Initial Report FDA Received Date01/03/2023
Supplement Report FDA Received Date02/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED; NONE REPORTED
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-